Standard Addition Calculator Online

Calculate the unknown concentration in a sample by single-point standard addition. Enter the sample volume, standard concentration, spike volume and the signals before and after.

Standard Addition Calculator

Free standard addition calculator that runs in your browser. Nothing is uploaded.

Volume-corrected

Assumes a linear response through zero. Use the same units for both volumes and for the result as for Cs.

How to use standard addition

1Measure the sample

Record the signal S₀ for your sample of volume Vx.

2Spike and measure again

Add volume Vs of standard Cs, mix and record S₁.

3Calculate

Enter all five values to get the concentration in the original sample.

Worked numbers

Cx = Cs·Vs·S₀ ÷ [S₁(Vx + Vs) − S₀·Vx]. This corrects for the small dilution caused by adding the spike.

Example values: Vx = 10 mL, Cs = 100 mg/L, Vs = 1 mL.

S₀S₁Cx (corrected)Without correction
20456.78 mg/L8.00 mg/L
203015.38 mg/L20.0 mg/L
508013.16 mg/L16.67 mg/L
10601.79 mg/L2.00 mg/L

When and why to use standard addition

This short guide explains what the standard addition calculator does, how to get the best result and when another tool is a better fit.

The matrix problem

In real samples, such as seawater, blood, soil extracts or food, other substances can boost or suppress the instrument signal. A calibration curve made in clean water would then give the wrong answer. Standard addition solves this by adding known amounts of the analyte to the sample itself, so the standard experiences exactly the same matrix effects as the unknown.

Standard Addition Calculator – S₀ to Cx
How the standard addition calculator works: S₀ in, Cx out.

It is common in atomic absorption, ICP, electrochemistry and fluorescence.

How the calculation works

If the signal is proportional to concentration, the increase after spiking tells you how much signal one unit of concentration produces in that matrix. Comparing that with the original signal gives the unknown. Adding the spike also dilutes the sample slightly, from Vx to Vx + Vs, which the formula corrects for. Ignoring this dilution can give a noticeably wrong result when the spike volume isn’t tiny.

The results also show the uncorrected value for comparison.

A worked example

An analyst measures lead in a 10 mL water sample and gets a signal of 20. After adding 1 mL of a 100 mg/L standard, the signal rises to 45. Cx = 100 × 1 × 20 ÷ (45 × 11 − 20 × 10) = 2,000 ÷ 295 = 6.78 mg/L. Without the volume correction the answer would be 8.00 mg/L, about 18% too high.

Report the result with the instrument’s uncertainty.

Getting reliable results

Choose a spike that roughly doubles or triples the signal, stay within the instrument’s linear range and subtract any blank signal before entering S₀ and S₁. Single-point addition is quick, but multiple spikes with a regression line are more robust and reveal non-linearity.

Check results against your lab’s procedures and product data sheets.

Related tools

To prepare the standards themselves, use the Calibration Standard Calculator. For ppm dilutions of stock solutions, use the PPM Dilution Calculator.

  • Spike the sample to cancel matrix effects
  • Correct for the spike’s dilution
  • Subtract the blank first

Further reading: The standard addition method. It explains the format behind this standard addition calculator in more detail.

Where it’s used

AAS and ICP

Metals in complex samples.

Electrochemistry

Ion-selective electrodes.

Food analysis

Minerals and additives.

Environmental

Water and soil extracts.

Quick fixes

“The spiked signal is too low…”

S₁ must rise enough above S₀ after dilution; check values and units.

Result looks too high

Subtract the blank signal from S₀ and S₁ first.

Units mixed

Use the same volume unit for Vx and Vs.

Non-linear response

Use several spikes and a regression line instead.

Frequently asked questions

What is standard addition?

Adding known analyte to the sample to measure it within its own matrix.

What formula is used?

Cx = Cs·Vs·S₀ ÷ [S₁(Vx+Vs) − S₀·Vx].

Why correct for volume?

The spike dilutes the sample slightly.

What units does Cx have?

The same as Cs.

How big should the spike be?

Enough to roughly double or triple the signal.

Should I subtract a blank?

Yes, before entering signals.

Is one spike enough?

It works, but several spikes are more reliable.

Does it need a linear response?

Yes, through zero.

Is my data saved?

No.

When is it better than a calibration curve?

When the sample matrix affects the signal.