When and why to use standard addition
This short guide explains what the standard addition calculator does, how to get the best result and when another tool is a better fit.
The matrix problem
In real samples, such as seawater, blood, soil extracts or food, other substances can boost or suppress the instrument signal. A calibration curve made in clean water would then give the wrong answer. Standard addition solves this by adding known amounts of the analyte to the sample itself, so the standard experiences exactly the same matrix effects as the unknown.
It is common in atomic absorption, ICP, electrochemistry and fluorescence.
How the calculation works
If the signal is proportional to concentration, the increase after spiking tells you how much signal one unit of concentration produces in that matrix. Comparing that with the original signal gives the unknown. Adding the spike also dilutes the sample slightly, from Vx to Vx + Vs, which the formula corrects for. Ignoring this dilution can give a noticeably wrong result when the spike volume isn’t tiny.
The results also show the uncorrected value for comparison.
A worked example
An analyst measures lead in a 10 mL water sample and gets a signal of 20. After adding 1 mL of a 100 mg/L standard, the signal rises to 45. Cx = 100 × 1 × 20 ÷ (45 × 11 − 20 × 10) = 2,000 ÷ 295 = 6.78 mg/L. Without the volume correction the answer would be 8.00 mg/L, about 18% too high.
Report the result with the instrument’s uncertainty.
Getting reliable results
Choose a spike that roughly doubles or triples the signal, stay within the instrument’s linear range and subtract any blank signal before entering S₀ and S₁. Single-point addition is quick, but multiple spikes with a regression line are more robust and reveal non-linearity.
Check results against your lab’s procedures and product data sheets.
Related tools
To prepare the standards themselves, use the Calibration Standard Calculator. For ppm dilutions of stock solutions, use the PPM Dilution Calculator.
- Spike the sample to cancel matrix effects
- Correct for the spike’s dilution
- Subtract the blank first
Further reading: The standard addition method. It explains the format behind this standard addition calculator in more detail.